Gastrostomy Tube Replacement


Article Author:
Rushikesh Shah


Article Editor:
Mili Shah


Editors In Chief:
Sebastiano Cassaro
Joseph Lee
Tanya Egodage


Managing Editors:
Avais Raja
Orawan Chaigasame
Carrie Smith
Abdul Waheed
Khalid Alsayouri
Frank Smeeks
Kristina Soman-Faulkner
Radia Jamil
Patrick Le
Sobhan Daneshfar
Anoosh Zafar Gondal
Saad Nazir
William Gossman
Pritesh Sheth
Hassam Zulfiqar
Navid Mahabadi
Steve Bhimji
John Shell
Matthew Varacallo
Heba Mahdy
Ahmad Malik
Mark Pellegrini
James Hughes
Beata Beatty
Nazia Sadiq
Hajira Basit
Phillip Hynes
Tehmina Warsi


Updated:
5/15/2019 9:55:14 PM

Introduction

A gastrostomy tube (G-tube) is indicated in patients with a need for long-term pre-pyloric feeding. Endoscopic gastrostomy tube placement is now the preferred modality over surgical gastrostomy owing to a less invasive nature and faster time to start feeding [1]. During endoscopic gastrostomy tube placement, a small incision is made over anterior abdominal wall through which tube is placed via push method (the tube inserted percutaneously through the incision) or pull technique (the tube is pulled out from the stomach through the incision). Once the tube is placed, the percutaneous tract starts maturing in about 1 to 2 weeks and usually well formed in 4 to 6 weeks. However, in certain conditions such as severe malnutrition, immunocompromised patients, and large ascites, tract maturity is further delayed. If a PEG tube is dislodged within a month after placement, then endoscopic replacement is recommended. However, if the tube is dislodged after 4 to 6 weeks when tract maturity is expected, bedside replacement is usually sufficient. If the tube is dislodged within 4 weeks of initial placement, patients are at significant risk of peritonitis and perforation due to peritoneal spillage of gastric contents through the immature track, and replacement should not be attempted without surgical consultation. Blind attempt to re-insert the tube or even foley placement in an immature tract can lead to inadvertent placement of the tube into the peritoneal cavity. If the tube needs to be replaced after the track is expected to be matured, will be discussed here in this topic.

Anatomy

During endoscopic placement of G-tube placement, a gastroscope is inserted, and air insufflation inflates the stomach. About 2.5 cm below the costal margin and medial to xiphoid process, an area of maximal transillumination is identified while shining an endoscope light through the anterior gastric wall transillumination. A good 1:1 contact is confirmed by firmly applying pressure on this point and visualizing stomach wall sliding off by this pressure at the same time. Once the site is confirmed, this area is marked with a surgical pen for incision site. After the tube placement with an endoscope, the mechanism of track maturation include anterior gastric wall adhering to the abdominal wall forming a gastrocutaneous fistula.

Indications

The indications of percutaneous endoscopic gastrostomy (PEG) tube replacement are:

  • Dislodged PEG tube: If the tube is dislodged within 24 hours, a temporary tube (e.g., Foley catheter) should be placed in the track to prevent track closure which starts occurring anytime between 8 to 24 hours and narrows further as time passes.
  • Malfunctioning PEG tube
  • Clogged G-tube not responding to traditional declogging measures

Contraindications

Bedside/bling PEG tube replacement is contraindicated in the following situations:

  • G-tube displacement of malfunction within 4 weeks of initial tube placement (due to track immaturity) [2].
  • Any signs of peritonitis
  • If the tube has been displaced for more than 24 hours, and the track has narrowed or closed.

Equipment

For bedside replacement of G-tube:

  • PEG tube (size similar to prior G-tube)
  • 10 mL saline syringe
  • Guidewire
  • Dressing kit
  • 2 x 2 and 4 x 4 gauge
  • Alcohol swabs and Povidone iodine swabs
  • Gastroscope (if endoscopic replacement is being considered)

Preparation

First, if the tube is still in place, it should be examined. There are cases of a malfunctioning tube. It is vital that the clinician take a careful history and review prior records when they are available, to know the type of tube (with internal mushroom cap versus internal balloon). The dressing is removed, and percutaneous incision site is examined for any signs of infection, track closure, or tube displacement in subcutaneous tissue. The area should be appropriately cleaned with sterilizing alcohol or povidone iodine. The area is covered with either sheath in PEG replacement kit or surgical drapes to avoid contaminating the surroundings during PEG replacement.

Technique

If PEG tube is not completely dislodged but has been clogged or malfunctioning, the old tube needs to be removed. If the old tube is endoscopically placed tube with internal mushroom tip, then gentle traction is applied to the external portion of the tube to remove internal mushroom. This has to be done with very gentle traction, and excessive force should be avoided at all cost as it can lead to disruption of an even well-formed track. If unusual resistance is felt during attempted removal of the tube by gentle traction, it is best to remove tube after endoscopic cutting of internal bumper/mushroom cap and removal of rest of the tube through external puling. If the old G-tube is with an internal balloon, the balloon is deflated by aspirating fluid through the balloon port which usually consists of 8 to 10 mL of fluid. After this step, once the tube is removed, the similar size PEG tube is prepared to be inserted.

For replacement G-tube, tubes with the balloon tip are usually preferred to minimize the risk of disruption of the old track. A similar diameter PEG tube as the prior tube should be used as a replacement tube. First, the balloon functioning is confirmed outside the body by injecting 8 to 10 mL of saline in balloon port. Once confirmed, the balloon is then deflated again. The balloon tip G-tube is either inserted into track blindly or via the use of guidewire with or without fluoroscopy and is advanced within the track. Once fully inside, the balloon is inflated with 8 to 10 mL of saline or distilled water. The tube is slowly retracted till gentle resistance is encountered due to the internal balloon opposing the gastric wall. At this point, the external portion of the tube is secured with dressing. Both ports of traditional G-tube (feeding port and meds port) are flushed and then capped. A dressing is then applied based on institutional policy. It is vital to apply protective dressing to avoid accidental pulling or dislodgement in patients with altered mentation.

The final step in PEG tube replacement is to confirm the placement. There are certain ways, gastric placement can be confirmed. The simplest method could be bedside method such as flushing air through the tube and auscultating for gastric gurgle. However, this method is quite unreliable. The most accurate tests are confirming internal balloon placement with an endoscope. However, this might not be cost-effective. The most commonly used technique in most practices is to obtain water-soluble contrast through the replacement tube and obtaining a post, contrast-enhanced, abdominal x-ray to confirm placement in the stomach [3].

Complications

The major complication of PEG tube replacement is the accidental peritoneal placement of the tube. Peritoneal tube placement should be suspected when after tube replacement or after resuming feeding through the tube, the patient starts having abdominal pain, fever, and signs of frank peritonitis on the exam. If this happens, a prompt investigation should be performed using contrast x-ray with water-soluble contrast or with CT scan of the abdomen. Tube feeds should be stopped immediately. Surgical consult should be obtained, and broad-spectrum antibiotics should be initiated as soon as possible. The patient is then managed conservatively with close hemodynamic monitoring along with surgical assistance if needed.

Other complications are rare, but the tube replacement site can get infected especially if the balloon is underinflated or not opposing the gastric wall. This can lead to leakage of gastric contents into the track. Bleeding due to trauma during tube replacement is rare but can happen.

Clinical Significance

G-tube malfunction is a commonly encountered in clinical practice both in inpatient and outpatient practice. At some institutes, G-tube replacement is exclusively done by either gastroenterology service or interventional radiology service depending on who initially placed the tube. However, if the patient is seen in the clinic, emergency room, or a nursing home for malfunctioning G-tube, an appropriate service, either gastroenterology or interventional radiology, should be consulted. If the tube has dislodged or come out, the gastrocutaneous fistula track will start to close as soon as in 8 hours but definitely in 24 hours. The longer the duration the tube has been out of place, higher the chances that the track has narrowed or closed. If more than 24 hours has elapsed after G-tube has been displaced, a blind forceful attempt to put a Foley or G-tube should not be done, and replacement should be referred to specialized services (gastrointestinal or interventional radiology). If G-tube has been dislodged for less than 24 hours, consider putting a temporary tube such as a Foley catheter (similar size as G-tube) to prevent the closure of the track and contact a specialty service for replacement. Once the tube is replaced, gastric placement confirmation with contrast-enhanced, abdominal x-ray is recommended [4]. If the patient develops any signs of peritonitis after tube replacement, accidental intraperitoneal tube placement should be suspected and further evaluated.

Enhancing Healthcare Team Outcomes

G-tube malfunction is commonly encountered by nurses, physician assistants, nurse practitioners and physicians in clinical practice. The team should have a working knowledge of how to handle G-tube problems and provide appropriate intervention and assistance in resolving the dysfunction.


Interested in Participating?

We are looking for contributors to author, edit, and peer review our vast library of review articles and multiple choice questions. In as little as 2-3 hours you can make a significant contribution to your specialty. In return for a small amount of your time, you will receive free access to all content and you will be published as an author or editor in eBooks, apps, online CME/CE activities, and an online Learning Management System for students, teachers, and program directors that allows access to review materials in over 500 specialties.

Improve Content - Become an Author or Editor

This is an academic project designed to provide inexpensive peer-reviewed Apps, eBooks, and very soon an online CME/CE system to help students identify weaknesses and improve knowledge. We would like you to consider being an author or editor. Please click here to learn more. Thank you for you for your interest, the StatPearls Publishing Editorial Team.

Gastrostomy Tube Replacement - Questions

Take a quiz of the questions on this article.

Take Quiz
An 85-year-old male is transferred from nursing home to emergency department as his G-tube came out accidentally about 3 hours ago. He is nonverbal secondary to a history of stroke with dysphagia and hemiplegia. The G-tube was placed by gastroenterology service 6 months ago. The prior tube site looks healthy and without any signs of infection or discharge. What should be done?



Click Your Answer Below


Would you like to access teaching points and more information on this topic?

Improve Content - Become an Author or Editor and get free access to the entire database, free eBooks, as well as free CME/CE as it becomes available. If interested, please click on "Sign Up" to register.

Purchase- Want immediate access to questions, answers, and teaching points? They can be purchased above at Apps and eBooks.


Sign Up
While cross covering for the medical floor at night, you are called to evaluate a patient as patient's G-tube is not functioning. Upon evaluation, the patient had G-tube placed by the gastroenterology service 1 week ago for dysphagia due to severe multiple sclerosis. The nurse confirmed patient was agitated and was pulling on G-tube despite repeated attempts to redirect him. On examination, you observe a mushroom-shaped button in subcutaneous tissue just underneath the skin. The area around G-tube site is extremely tender with rebound. His pulse is 130, Temperature 38.3 and he is normotensive. What would be your next step?



Click Your Answer Below


Would you like to access teaching points and more information on this topic?

Improve Content - Become an Author or Editor and get free access to the entire database, free eBooks, as well as free CME/CE as it becomes available. If interested, please click on "Sign Up" to register.

Purchase- Want immediate access to questions, answers, and teaching points? They can be purchased above at Apps and eBooks.


Sign Up
You are seeing a patient with dislodged G-tube. The tube was placed 4 years ago for stroke-related dysphagia by the gastroenterology service. The tube has been replaced 6 times in last 4 years due to accidental dislodgement. You consulted gastroenterology service who came and replaced G-tube at the bedside and asked to confirm placement in the stomach prior to use. What is the preferred and most cost-effective way to confirm placement in the stomach?



Click Your Answer Below


Would you like to access teaching points and more information on this topic?

Improve Content - Become an Author or Editor and get free access to the entire database, free eBooks, as well as free CME/CE as it becomes available. If interested, please click on "Sign Up" to register.

Purchase- Want immediate access to questions, answers, and teaching points? They can be purchased above at Apps and eBooks.


Sign Up
A nursing home resident is evaluated for a PEG tube site infection. On evaluation, G-tube was placed 6 months ago. The track is dilated and a thick discharge around the tube. The tube itself looks fine and flushes. The skin around the tube is red and warm. What is the next step?



Click Your Answer Below


Would you like to access teaching points and more information on this topic?

Improve Content - Become an Author or Editor and get free access to the entire database, free eBooks, as well as free CME/CE as it becomes available. If interested, please click on "Sign Up" to register.

Purchase- Want immediate access to questions, answers, and teaching points? They can be purchased above at Apps and eBooks.


Sign Up
You are seeing a patient in your clinic who came as he is having pain at the G-tube site. The G-tube was placed 4 months ago for dysphagia. He states the tube has not worked for the last 4 days and he is having worsening pain at the G-tube site. You are unable to flush the tube. The external bumper is jammed into the abdominal wall and you unable to rotate the tube with the bumper. What is the most likely diagnosis?



Click Your Answer Below


Would you like to access teaching points and more information on this topic?

Improve Content - Become an Author or Editor and get free access to the entire database, free eBooks, as well as free CME/CE as it becomes available. If interested, please click on "Sign Up" to register.

Purchase- Want immediate access to questions, answers, and teaching points? They can be purchased above at Apps and eBooks.


Sign Up

Gastrostomy Tube Replacement - References

References

Gastrostomy tube placement outcomes: comparison of surgical, endoscopic, and laparoscopic methods., Bankhead RR,Fisher CA,Rolandelli RH,, Nutrition in clinical practice : official publication of the American Society for Parenteral and Enteral Nutrition, 2005 Dec     [PubMed]
Evaluation of gastropexy and stoma tract maturation using a novel introducer kit for percutaneous gastrostomy in a porcine model., Maxwell CI,Hilden K,Glasgow RE,Ollerenshaw J,Carlisle JG,Fang JC,, JPEN. Journal of parenteral and enteral nutrition, 2011 Sep     [PubMed]
Percutaneous feeding tube replacement in the ED--are confirmatory x-rays necessary?, Jacobson G,Brokish PA,Wrenn K,, The American journal of emergency medicine, 2009 Jun     [PubMed]
Gastrostomy tube replacement in a pediatric ED: frequency of complications and impact of confirmatory imaging., Showalter CD,Kerrey B,Spellman-Kennebeck S,Timm N,, The American journal of emergency medicine, 2012 Oct     [PubMed]

Disclaimer

The intent of StatPearls is to provide practice questions and explanations to assist you in identifying and resolving knowledge deficits. These questions and explanations are not intended to be a source of the knowledge base of all of medicine, nor is it intended to be a board or certification review of Surgery-General. The authors or editors do not warrant the information is complete or accurate. The reader is encouraged to verify each answer and explanation in several references. All drug indications and dosages should be verified before administration.

StatPearls offers the most comprehensive database of free multiple-choice questions with explanations and short review chapters ever developed. This system helps physicians, medical students, dentists, nurses, pharmacists, and allied health professionals identify education deficits and learn new concepts. StatPearls is not a board or certification review system for Surgery-General, it is a learning system that you can use to help improve your knowledge base of medicine for life-long learning. StatPearls will help you identify your weaknesses so that when you are ready to study for a board or certification exam in Surgery-General, you will already be prepared.

Our content is updated continuously through a multi-step peer review process that will help you be prepared and review for a thorough knowledge of Surgery-General. When it is time for the Surgery-General board and certification exam, you will already be ready. Besides online study quizzes, we also publish our peer-reviewed content in eBooks and mobile Apps. We also offer inexpensive CME/CE, so our content can be used to attain education credits while you study Surgery-General.